Real-time 3-D ultrasound imaging using sparse synthetic aperture beamforming
Author :
Lockwood, Geoffrey R. ; Talman, James R. ; Brunke, Shelby S.
Author_Institution :
Dept. of Biomed. Eng., Cleveland Clinic Found., OH, USA
Volume :
45
Issue :
4
fYear :
1998
fDate :
7/1/1998 12:00:00 AM
Firstpage :
980
Lastpage :
988
Abstract :
A method for real-time three-dimensional (3-D) ultrasound imaging using a mechanically scanned linear phased array is proposed. The high frame rate necessary for real-time volumetric imaging is achieved using a sparse synthetic aperture beamforming technique utilizing only a few transmit pulses for each image. Grating lobes in the two-way radiation pattern are avoided by adjusting the transmit element spacing and the receive aperture functions to account for the missing transmit elements. The signal loss associated with fewer transmit pulses is minimized by increasing the power delivered to each transmit element and by using multiple transmit elements for each transmit pulse. By mechanically rocking the array, in a way similar to what is done with an annular array, a 3-D set of images can be collected in the time normally required for a single image.
Keywords :
acoustic signal processing; array signal processing; ultrasonic arrays; ultrasonic imaging; grating lobe; linear phased array; mechanical scanning; radiation pattern; real-time three-dimensional ultrasound imaging; sparse synthetic aperture beamforming; volumetric imaging; Apertures; Array signal processing; Gratings; Phased arrays; Propagation losses; Ultrasonic imaging;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on