Title :
Super-harmonic imaging: development of an interleaved phased-array transducer
Author :
Van Neer, Paul L M J ; Matte, Guillaume ; Danilouchkine, Mikhail G. ; Prins, Christian ; Van Den Adel, F. ; de Jong, Nico
Author_Institution :
Dept. of Biomed. Eng., Erasmus Med. Centre, Rotterdam, Netherlands
fDate :
2/1/2010 12:00:00 AM
Abstract :
For several years, the standard in ultrasound imaging has been second-harmonic imaging. A new imaging technique dubbed "super-harmonic imaging" (SHI) was recently proposed. It takes advantage of the higher-third to fifth-harmonics arising from nonlinear propagation or ultrasound-contrast-agent (UCA) response. Next to its better suppression of near-field artifacts, tissue SHI is expected to improve axial and lateral resolutions resulting in clearer images than second-harmonic imaging. When SHI is used in combination with UCAs, a better contrast-to-tissue ratio can be obtained. The use of SHI implies a large dynamic range and requires a sufficiently sensitive array over a frequency range from the transmission frequency up to its fifth harmonic (bandwidth > 130%). In this paper, we present the characteristics and performance of a new interleaved dual frequency array built chiefly for SHI. We report the rationale behind the design choice, frequencies, aperture, and piezomaterials used. The array is efficient both in transmission and reception with well-behaved transfer functions and a combined -6-dB bandwidth of 144%. In addition, there is virtually no contamination of the harmonic components by spurious transducer transmission, due to low element-to-element crosstalk (< 30 dB) and a low transmission efficiency of the odd harmonics (< 46 dB). The interleaved array presented in this article possesses ideal characteristics for SHI and is suitable for other methods like second-harmonic, subharmonic, and second-order ultrasound field (SURF) imaging.
Keywords :
acoustic signal processing; biomedical transducers; biomedical ultrasonics; harmonic generation; nonlinear acoustics; piezoelectric transducers; ultrasonic imaging; ultrasonic transducer arrays; contrast to tissue ratio; fifth harmonics; fourth harmonics; inter-element crosstalk; interleaved dual frequency array; interleaved phased array transducer; near field artifact suppression; nonlinear ultrasound propagation; piezomaterials; superharmonic imaging; third harmonics; tissue SHI; transfer functions; ultrasound contrast agent response; ultrasound imaging; Apertures; Bandwidth; Contamination; Dynamic range; Frequency; High-resolution imaging; Image resolution; Transfer functions; Ultrasonic imaging; Ultrasonic transducers; Computer Simulation; Contrast Media; Echocardiography, Doppler; Nonlinear Dynamics; Normal Distribution; Transducers;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
DOI :
10.1109/TUFFC.2010.1426