DocumentCode :
1424127
Title :
Series resonant ZCS-PFM DC-DC converter with multistage rectified voltage multiplier and dual-mode PFM control scheme for medical-use high-voltage X-ray power generator
Author :
Sun, J. ; Ding, X. ; Nakaoka, M. ; Takano, H.
Author_Institution :
ESSD, APC Japan Ltd., Tokyo, Japan
Volume :
147
Issue :
6
fYear :
2000
fDate :
11/1/2000 12:00:00 AM
Firstpage :
527
Lastpage :
534
Abstract :
In general, high-voltage DC power supplies employing a variety of high-frequency inverters are implemented for constant value control schemes. In particular, their good transient and steady state performances cannot be achieved under wide load variations for a medical-use high-voltage X-ray generator, ranging from 20 kV to 150 kV in the output voltage and from 0.5 mA to 1250 mA, respectively. A high-voltage DC power supply designed for X-ray power generator applications is considered, which uses a series resonant inverter-linked multistage DC voltage multiplier instead of a conventional high-voltage diode module rectifier connected to the secondary side of a high-voltage transformer. A constant on-time/variable frequency control scheme of this converter operating at zero-current soft switching mode is described. Introducing the capacitor-diode cascaded multistage voltage multiplier, the secondary turn numbers and secondary-side stray capacitance of the high-voltage, high-frequency transformer, as well as the rectifier diode voltage ratings, can be greatly reduced. It is shown that the proposed converter control scheme of the two-step selective changed frequency selection switching is more effective for improving the output voltage responses. The series resonant high frequency transformer-linked voltage-multiplying rectifier is evaluated for an X-ray high-voltage generator on the basis of simulation analysis and observed data in experiment
Keywords :
DC-DC power convertors; X-ray production; biomedical equipment; frequency control; high-frequency transformers; invertors; power supplies to apparatus; pulse frequency modulation; resonant power convertors; switching circuits; voltage control; voltage multipliers; 0.5 to 1250 mA; 20 to 150 kV; X-ray high-voltage generator; capacitor-diode cascaded multistage voltage multiplier; constant on-time/variable frequency control; constant value control schemes; dual-mode PFM control scheme; high-frequency inverters; high-voltage DC power supply; high-voltage high-frequency transformer; high-voltage transformer secondary side; medical-use high-voltage X-ray generator; medical-use high-voltage X-ray power generator; multistage rectified voltage multiplier; output voltage control; output voltage responses improvement; rectifier diode voltage ratings; secondary turn numbers; secondary-side stray capacitance; series resonant ZCS-PFM DC-DC converter; series resonant high frequency transformer; series resonant inverter-linked multistage DC voltage multiplier; simulation analysis; steady state performance; transient performance; two-step selective changed frequency selection switching; wide load variations; zero-current soft switching mode;
fLanguage :
English
Journal_Title :
Electric Power Applications, IEE Proceedings -
Publisher :
iet
ISSN :
1350-2352
Type :
jour
DOI :
10.1049/ip-epa:20000711