Title :
Energy-dependent systematic errors in dual-energy X-ray CT
Author :
Goh, K.L. ; Liew, S.C. ; Hasegawa, B.H.
Author_Institution :
Dept. of Phys., Nat. Univ. of Singapore, Singapore
fDate :
4/1/1997 12:00:00 AM
Abstract :
Dual-energy X-ray computed tomography (DECT) is a technique which is designed to allow the determination of energy-independent material properties. In this study, results of a computer simulation show that energy-dependent systematic errors exist in the values of attenuation coefficients synthesized using the basis material decomposition technique with acrylic and aluminum as the basis materials, especially when a high atomic number element such as iodine (e.g., from radiographic contrast media) is present in the body. The errors are reduced when an acrylic and an iodine-water mixture are used as the basis materials. The authors propose a simple theoretical model for the calculation of energy-dependent systematic errors using effective energies at the lower and higher energy windows of the X-ray spectrum used in the DECT system. The calculated errors agree well with the errors observed in the simulation. These results suggest that the observed systematic errors are predominantly due to the energy dependence of the basis material coefficients
Keywords :
X-ray absorption; computerised tomography; digital simulation; measurement errors; Al; I; X-ray spectrum; acrylic; attenuation coefficients; basis material coefficients; computer simulation; dual-energy X-ray CT; effective energies; energy-dependent systematic errors; energy-independent material properties; high atomic number element; iodine-water mixture; medical diagnostic imaging; radiographic contrast media; simple theoretical model; Aluminum; Atomic measurements; Attenuation; Computed tomography; Computer errors; Computer simulation; Image reconstruction; Material properties; Radiography; X-ray imaging;
Journal_Title :
Nuclear Science, IEEE Transactions on