    Improved high-intensity focused ultrasound (HIFU) surgical applicators are required for use in a surgical environment. We report on the performance and characteristics of a new solid-cone HIFU applicator. Previous HIFU devices used a water-filled stand-off to couple the ultrasonic energy from the transducer to the treatment area. The new applicator uses a spherically-focused element and a solid aluminum cone to guide and couple the ultrasound to the tissue. Compared with the water-filled applicators, this new applicator is simpler to set up and manipulate, cannot leak, prevents the possibility of cavitation within the coupling device, and is much easier to sterilize and maintain during surgery. The design minimizes losses caused by shear wave conversion found in tapered solid acoustic velocity transformers operated at high frequencies. Computer simulations predicted good transfer of longitudinal waves. Impedance measurements, beam plots, Schlieren images, and force balance measurements verified strong focusing and suitable transfer of acoustic energy into water. At the focus, the -3 dB beam dimensions are 1.2 mm (axial)×0.3 mm (transverse). Radiation force balance measurements indicate a power transfer efficiency of 40%. In vitro and in vivo tissue experiments confirmed the applicator´s ability to produce hemostasis.
  • Keywords
    biomedical transducers; biomedical ultrasonics; surgery; ultrasonic transducers; 40 percent; Schlieren images; cavitation; hemostasis; high intensity focused ultrasound applicator; impedance measurements; longitudinal waves; power transfer efficiency; radiation force balance measurements; shear wave conversion; solid-cone HIFU applicator; spherically-focused element; sterilizing; surgical applications; Acoustic beams; Acoustic measurements; Applicators; Focusing; Force measurement; Impedance measurement; Solids; Surgery; Ultrasonic imaging; Ultrasonic transducers; Acoustics; Animals; Blood Vessels; Computer Simulation; Electric Impedance; Hemostasis, Surgical; Liver; Models, Biological; Rabbits; Spleen; Transducers; Ultrasonic Therapy; Vascular Surgical Procedures;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.895907
  • Filename
    895907