    A New Indirect Calibration Method for an Equivalent Source Mismatch of a Power Splitter Using Measurements at Only One Port

  • Author

    Suzuki, Kiyoshi

  • Author_Institution
    Meas. Stand. Center, Agilent Technol. Int. Japan, Ltd., Tokyo, Japan
  • Volume
    60
  • Issue
    7
  • fYear
    2011
  • fDate
    7/1/2011 12:00:00 AM
  • Firstpage
    2684
  • Lastpage
    2689
  • Abstract
    A new measurement procedure is described for determining the equivalent source mismatch ΓG of a power splitter. It uses a vector network analyzer (VNA) configured for one-port calibration and measurement. The power splitter ports that are not connected to the VNA test port are loaded in turn with characterized standards, such as an open, a short, a load, and a termination. It also takes advantage of the reciprocity characteristic of the power splitter. As the system impedance is normalized to the impedance of the termination, the termination works as a perfect load in measurements and calculations for the S-parameters of the power splitter. ΓG is calculated from the renormalized S-parameters to Z0 . The propagation of uncertainties is analyzed.
  • Keywords
    S-parameters; calibration; electric sensing devices; network analysers; power measurement; S-parameter; equivalent source mismatch; indirect calibration method; power measurement; power sensor; power splitter; vector network analyzer; Calibration; Loading; Power measurement; Scattering parameters; Sensitivity; Standards; Uncertainty; Calibration; equivalent source mismatch; power sensor; power splitter; uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2010.2096891
  • Filename
    5688454