Abstract :
The problem of providing surgical navigation using image overlays on the operative scene can be split into four main tasks-calibration of the optical system; registration of preoperative images to the patient; system and patient tracking, and display using a suitable visualization scheme. To achieve a convincing result in the magnified microscope view a very high alignment accuracy is required. The authors have simulated an entire image overlay system to establish the most significant sources of error and improved each of the stages involved. The microscope calibration process has been automated. The authors have introduced bone-implanted markers for registration and incorporated a locking acrylic dental stent (LADS) for patient tracking. The LADS can also provide a less-invasive registration device with mean target error of 0.7 mm in volunteer experiments. These improvements have significantly increased the alignment accuracy of the authors´ overlays. Phantom accuracy is 0.3-0.5 mm and clinical overlay errors were 0.5-1.0 mm on the bone fiducials and 0.5-4 mm on target structures. The authors have improved the graphical representation of the stereo overlays. The resulting system provides three-dimensional surgical navigation for microscope-issisted guided interventions (MAGI).
Keywords :
biomedical equipment; calibration; dentistry; image registration; medical image processing; optical microscopes; orthopaedics; surgery; alignment accuracy; bone fiducials; bone-implanted markers; clinical overlay errors; locking acrylic dental stent; medical instrumentation; microscope-assisted guided interventions system; optical system calibration; patient tracking; phantom accuracy; surgical navigation; target error; target structures; volunteer experiments; Calibration; Dentistry; Displays; Imaging phantoms; Layout; Microscopy; Navigation; Surgery; Target tracking; Visualization; Computer Simulation; Equipment Design; Humans; Microscopy; Surgical Procedures, Operative;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.896784
Filename :
896784
Link To Document :
بازگشت