• DocumentCode
    1430330
  • Title

    An enhanced method for characterizing successive approximation converters

  • Author

    Young, Chris A.

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Wales Univ., Swansea, UK
  • Volume
    39
  • Issue
    2
  • fYear
    1990
  • fDate
    4/1/1990 12:00:00 AM
  • Firstpage
    335
  • Lastpage
    339
  • Abstract
    A strategy for characterizing successive approximation, analog-to-digital converters with quad current switch architectures is described. It permits detailed parameter specification (accuracy and linearity) without recourse to testing every transition voltage. The results show an encouraging reduction in the residual error from 0.144 step (from a currently used test) to 0.048 step with the proposed test for the specific case of 12-b converters. To achieve this result, only 2.6% of the transition voltages require measurement. The percentage of codes to be tested reduces dramatically as the number of bits of the converter increases. The evolution of the proposed test is investigated by examining the cause and effect of superposition errors
  • Keywords
    analogue-digital conversion; characteristics measurement; electronic equipment testing; measurement errors; accuracy; analog-to-digital converters; linearity; parameter specification; quad current switch architectures; residual error; static transfer characteristic; successive approximation converters; superposition errors; Analog-digital conversion; Automatic testing; Helium; Industrial control; Linearity; Switches; Switching converters; Time measurement; Transfer functions; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.52511
  • Filename
    52511