• DocumentCode
    1430980
  • Title

    Axial Anisotropic Conductivity Imaging Based on Projected Current Density in MREIT

  • Author

    Nam, Hyun Soo ; Kwon, Oh In

  • Author_Institution
    Dept. of Math., Konkuk Univ., Seoul, South Korea
  • Volume
    29
  • Issue
    3
  • fYear
    2010
  • fDate
    3/1/2010 12:00:00 AM
  • Firstpage
    781
  • Lastpage
    789
  • Abstract
    An aim of magnetic resonance electrical impedance tomography (MREIT) is to visualize the internal current density and/or the conductivity of an imaging object. In MREIT, it is desirable to use just one component of the internal magnetic flux density vector B=(B x,B y,B z) caused by the injected current, measured without rotating the object. We present a method of visualizing the axial anisotropic conductivity tensor by use of the measured magnetic flux density B z data. The method involves the use of a projected current density, which is a uniquely and stably determined component of the internal current generated by the injected current, derived from the measured B z data. Each component of the axial anisotropic conductivity is recovered by matching the measured B z data with a determined intermediate isotropic conductivity and the projected currents. Results from numerical simulations demonstrate that the proposed algorithm is robust to noise and stably determines the anisotropic conductivity tensor on the imaging slice. For a practical implementation, we studied a postmortem canine brain case to visualize each component of the anisotropic conductivity. We observed that the reconstructed anisotropic conductivity images clearly reflects the anisotropic property of the white matter in the direction parallel to its fibers.
  • Keywords
    bioelectric phenomena; biomedical MRI; brain; computerised tomography; current density; magnetic flux; MREIT; axial anisotropic conductivity imaging; axial anisotropic conductivity tensor; injected current; intermediate isotropic conductivity; internal current density; internal magnetic flux density vector; magnetic resonance electrical impedance tomography; postmortem canine brain; projected current density; Anisotropic magnetoresistance; Conductivity measurement; Current density; Current measurement; Data visualization; Density measurement; Magnetic flux density; Magnetic resonance; Magnetic resonance imaging; Tensile stress; Axial anisotropic conductivity; current density; magnetic resonance electrical impedance tomography (MREIT); magnetic resonance imaging (MRI); Animals; Anisotropy; Brain; Computer Simulation; Dogs; Electric Impedance; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Phantoms, Imaging;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2009.2036440
  • Filename
    5423311