• DocumentCode
    1438833
  • Title

    Range-uncertainty imaging for obtaining dose perturbations in proton therapy

  • Author

    Schneider, Uwe ; Tourovsky, Alexander

  • Author_Institution
    Div. of Med. Phys., City Hosp. Triemli, Zurich, Switzerland
  • Volume
    45
  • Issue
    5
  • fYear
    1998
  • fDate
    10/1/1998 12:00:00 AM
  • Firstpage
    2309
  • Lastpage
    2313
  • Abstract
    The proton dose distribution of a monoenergetic parallel proton beam in a biological sample (a sheep head in a mold) has been calculated using a Monte Carlo code. The calculations have been performed both including the effect of multiple Coulomb scattering of the protons at the atoms in the phantom and without it. The resulting dose distributions were subtracted from each other and their absolute value was integrated in the direction of the incident proton beam. This projected dose difference was compared to measurements of proton range uncertainties obtained using proton radiography data. A nearly linear relationship between the dose perturbations and the range variations was found which indicates that measured range variations can be used to predict dose perturbations in proton therapy
  • Keywords
    Monte Carlo methods; biological effects of ionising particles; dosimetry; energy loss of particles; medical computing; perturbation theory; proton effects; radiation therapy; Monte Carlo code; biological sample; dose perturbations; linear relationship; monoenergetic parallel proton beam; multiple Coulomb scattering; projected dose difference; proton dose distribution; proton radiography; proton therapy; range variations; range-uncertainty imaging; sheep head; Atomic measurements; Biological information theory; Head; Imaging phantoms; Medical treatment; Monte Carlo methods; Particle beams; Protons; Radiography; Scattering;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.725269
  • Filename
    725269