Abstract :
A system that allows detection of single and double differential surface myoelectric signals during electrical stimulation of either the nerve or the muscle motor point, using surface electrodes, is described. The input referred stimulation artifact is limited to less than 24 mu V pp by a switching constant current/constant voltage output stage, slew rate limiting, and time windowing of the detected signal. The system is suitable for muscle crosstalk and for muscle fatigue investigation during functional electrical stimulation, in addition to measuring muscle or nerve conduction velocity.
Keywords :
bioelectric potentials; biological techniques and instruments; muscle; electrical stimulation; muscle crosstalk; muscle fatigue; muscle motor point; myoelectric-evoked potential recordings; nerve conduction velocity; stimulation artifacts suppression; surface electrodes; time windowing; Crosstalk; Electric variables measurement; Electrical stimulation; Electrodes; Fatigue; Muscles; Neuromuscular stimulation; Signal detection; Velocity measurement; Voltage; Electric Stimulation; Electrodes; Electromyography; Equipment Design; Humans; Muscles; Neural Conduction;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.7278
Filename :
7278
Link To Document :
بازگشت