DocumentCode
1441240
Title
Suppression of stimulation artifacts from myoelectric-evoked potential recordings
Author
Knaflitz, Marco ; Merletti, Roberto
Author_Institution
NeuroMuscular Res. Center, Boston Univ., MA, USA
Volume
35
Issue
9
fYear
1988
Firstpage
758
Lastpage
763
Abstract
A system that allows detection of single and double differential surface myoelectric signals during electrical stimulation of either the nerve or the muscle motor point, using surface electrodes, is described. The input referred stimulation artifact is limited to less than 24 mu V pp by a switching constant current/constant voltage output stage, slew rate limiting, and time windowing of the detected signal. The system is suitable for muscle crosstalk and for muscle fatigue investigation during functional electrical stimulation, in addition to measuring muscle or nerve conduction velocity.
Keywords
bioelectric potentials; biological techniques and instruments; muscle; electrical stimulation; muscle crosstalk; muscle fatigue; muscle motor point; myoelectric-evoked potential recordings; nerve conduction velocity; stimulation artifacts suppression; surface electrodes; time windowing; Crosstalk; Electric variables measurement; Electrical stimulation; Electrodes; Fatigue; Muscles; Neuromuscular stimulation; Signal detection; Velocity measurement; Voltage; Electric Stimulation; Electrodes; Electromyography; Equipment Design; Humans; Muscles; Neural Conduction;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.7278
Filename
7278
Link To Document