Title :
Combined detectors based on ZnSe(Te), CsI(Tl) and Si-PIN-PD for separate detection of alpha, beta and gamma radiation
Author :
Ryzhikov, V. ; Gal´chinetski, L. ; Galkin, S. ; Danshin, E. ; Kvitnitskaya, V. ; Silin, V. ; Chernikov, V.
Author_Institution :
STC for Radiat. Instrum. Concern Inst. for Single Crystals, Kharkov, Ukraine
fDate :
12/1/2000 12:00:00 AM
Abstract :
Combined detectors (CD) comprising ZnSe(Te), CsI(Tl), and Si-PIN-PD can be used for detection of charged particles and gamma quanta. The detection is based on selective sensitivity and the different decay times of these scintillators for these kinds of radiations. In this paper, we outline possibilities and limitations of CD in which signals are read by an 18×18 mm2 Hamamatsu S3204 Si-PIN-PD placed on the back side of CsI(Tl) crystal. A specially designed charge-sensitive preamplifier connected to a PD has equivalent noise level of 700 electrons during shaping time of 5 μs. The CD-Si-PIN-PD system is compatible with modern computers
Keywords :
alpha-particle detection; beta-ray detection; caesium compounds; gamma-ray detection; nuclear electronics; preamplifiers; silicon radiation detectors; solid scintillation detectors; tellurium; thallium; zinc compounds; CsI(Tl); CsI:Tl; Si; Si-PIN-PD; ZnSe(Te); ZnSe:Te; alpha radiation; beta radiation; charge-sensitive preamplifier; combined detectors; gamma radiation; selective sensitivity; shaping time; Electrons; Gamma ray detection; Gamma ray detectors; Ionizing radiation; Neutrons; Photonic crystals; Preamplifiers; Radiation detectors; Semiconductor radiation detectors; Solid scintillation detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on