Title :
A miniature SQUID magnetometer with direct read-out electronics
Author :
Polushkin, V. ; Wallis, M. ; Glowacka, D. ; Matthews, A. ; Lumley, J.M.
Author_Institution :
Sci Res. Div., Oxford Instrum., Cambridge, UK
fDate :
6/1/1997 12:00:00 AM
Abstract :
A novel miniature SQUID magnetometer with direct read-out electronics has been developed. A high flux-to-voltage transfer factor of up to 2500 mV//spl Phi//sub 0/ is achieved without additional positive feedback (/spl Phi//sub 0/ is the flux quantum). A flux resolution of 8/spl times/10/sup -7/ /spl Phi//sub 0//Hz/sup 1/2 /, corresponding to a magnetic moment sensitivity of approximately 2/spl times/10/sup -2/0 Am/sup 2//Hz/sup 1/2 / has been measured. This magnetometer can be used to study the dynamic magnetization properties (including NMR and MRI) of micron and sub-micron size particles and for high-resolution surface magnetometry.
Keywords :
SQUID magnetometers; MRI; NMR; direct read-out electronics; dynamic magnetization; flux resolution; flux-to-voltage transfer factor; magnetic moment sensitivity; miniature SQUID magnetometer; sub-micron particles; surface magnetometry; Bandwidth; Energy resolution; Feedback loop; Magnetic flux; Magnetic noise; Magnetic semiconductors; Magnetic sensors; Preamplifiers; SQUID magnetometers; Superconducting magnets;
Journal_Title :
Applied Superconductivity, IEEE Transactions on