• DocumentCode
    1449541
  • Title

    Axicon: a new laser beam delivery system for corneal surgery

  • Author

    Ren, Qiushi ; Birngruber, Reginald

  • Author_Institution
    Wellman Labs. of Photomed., Massachusetts Gen. Hospital, Boston, MA, USA
  • Volume
    26
  • Issue
    12
  • fYear
    1990
  • fDate
    12/1/1990 12:00:00 AM
  • Firstpage
    2305
  • Lastpage
    2308
  • Abstract
    A new laser beam delivery system using axicon beam shaping has been designed to produce controllable surface ablation as well as trephination for the corneal surgery. Noncontact cornea trephination was performed on human eye-bank eyes by using a Q-switched Er:YAG laser with a axicon-lens combination. The results shows that the uniformity of the trephination along the ring depends on the radial symmetry of the beam. The histological results of the samples show that the damage zone of the laser cut is less than 10 μm. This demonstrates the possibility of replacing mechanical trephination in corneal surgery
  • Keywords
    biomedical equipment; eye; laser applications in medicine; surgery; Q-switched Er:YAG laser; YAG:Er; YAl5O12:Er; axicon beam shaping; axicon-lens combination; controllable surface ablation; corneal surgery; damage zone; histological results; human eye-bank eyes; laser beam delivery system; laser cut; noncontact cornea trephination; radial symmetry; ring; Control systems; Laser ablation; Laser beam cutting; Laser beams; Laser surgery; Optical control; Optical design; Ring lasers; Shape control; Surface emitting lasers;
  • fLanguage
    English
  • Journal_Title
    Quantum Electronics, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    0018-9197
  • Type

    jour

  • DOI
    10.1109/3.64369
  • Filename
    64369