DocumentCode :
1453027
Title :
Laparoscopic placement of electrodes for diaphragm pacing using stimulation to locate the phrenic nerve motor points
Author :
Schmit, Brian D. ; Stellato, Thomas A. ; Miller, Michael E. ; Mortimer, J. Thomas
Author_Institution :
VA Med. Center, Cleveland, OH, USA
Volume :
6
Issue :
4
fYear :
1998
fDate :
12/1/1998 12:00:00 AM
Firstpage :
382
Lastpage :
390
Abstract :
Laparoscopic mapping of the phrenic nerve motor points using test stimulation was conducted for the implant of epimysial electrodes for diaphragm pacing in dogs. Both visual assessment of muscle activation and measurements of recruitment were useful for identifying an implant location resulting in a mean electrode placement approximately 14 mm from the phrenic nerve motor points in 16 dogs. Postmortem analysis of the stimulus test site locations and corresponding recruitment curves suggested that the phrenic nerve motor points could be predicted during the laparoscopic procedure to within 4.5 mm of the anatomical motor point
Keywords :
bioelectric phenomena; biomedical electrodes; neuromuscular stimulation; pneumodynamics; surgery; 14 mm; 4.5 mm; anatomical motor point; diaphragm pacing; dogs; electrical stimulation; epimysial electrodes implant; mean electrode placement; muscle activation; phrenic nerve motor points location; recruitment curves; ventilation; Abdomen; Biomedical measurements; Dogs; Electrical stimulation; Electrodes; Implants; Laparoscopes; Recruitment; Surgery; Testing;
fLanguage :
English
Journal_Title :
Rehabilitation Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1063-6528
Type :
jour
DOI :
10.1109/86.736152
Filename :
736152
Link To Document :
بازگشت