Abstract :
The goal of this study is to develop a focal zone sharpening strategy that produces more precise lesions for pulsed cavitational ultrasound therapy, or histotripsy. Precise and well-confined lesions were produced by locally suppressing cavitation in the periphery of the treatment focus without affecting cavitation in the center. The local suppression of cavitation was achieved using cavitation nuclei preconditioning pulses to actively control cavitation in the periphery of the focus. A 1-MHz 513-element therapeutic array was used to generate both the therapy and the nuclei preconditioning pulses. For therapy, 10-cycle bursts at 100-Hz pulse repetition frequency with P-/P+ pressure of 21/76 MPa were delivered to the geometric focus of the therapeutic array. For nuclei preconditioning, a different pulse was delivered to an annular region immediately surrounding the focus before each therapy pulse. A parametric study on the effective pressure, pulse duration, and delivery time of the preconditioning pulse was conducted in red blood cell-gel phantoms, where cavitational damage was indicated by the color change resulting from local cell lysis. Results showed that a short-duration (20 μs) preconditioning pulse at a medium pressure (P-/P+ pressure of 7.2/13.6 MPa) delivered shortly before (30 μs) the therapy pulse substantially suppressed the peripheral damage by 77 ± 13% while complete fractionation in the focal center was maintained. High-speed imaging of the bubble cloud showed a substantial decrease in the maximum width of the bubble cloud by 48 ± 24% using focal zone sharpening. Experiments in ex vivo livers confirmed that highly confined lesions were produced in real tissues as well as in the phantoms. This study demonstrated the feasibility of active focal zone sharpening using cavitation nuclei preconditioning, allowing for increased treatment precision compared with the natural focal width of the therapy transducer.
Keywords :
biomedical imaging; biomedical ultrasonics; blood; bubbles; cavitation; cellular biophysics; fractionation; gels; phantoms; radiation therapy; active focal zone sharpening; bubble; color change; effective pressure; fractionation; frequency 1 MHz; frequency 100 Hz; high-precision treatment; high-speed imaging; histotripsy; lesions; local cell lysis; nuclei preconditioning pulses; pressure 7.2 MPa to 76 MPa; pulse duration; pulsed cavitational ultrasound therapy; red blood cell-gel phantoms; therapeutic array; Arrays; Laser beams; Lesions; Medical treatment; Phantoms; Pixel; Ultrasonic imaging; Ablation Techniques; Analysis of Variance; Animals; Dogs; Equipment Design; Erythrocytes; Histological Techniques; Liver; Phantoms, Imaging; Photography; Ultrasonic Therapy;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2011.1808
Filename :
5716448
Link To Document :
بازگشت