DocumentCode
1455488
Title
Comparison of the Josephson voltage standards of the SP and the BIPM
Author
Reymann, Dominique ; Witt, Thomas J. ; Eklund, Gunnar ; Pajander, Hannu ; Nilsson, Håkan
Author_Institution
Bur. Int. des Poids et Mesures, Sevres, France
Volume
46
Issue
2
fYear
1997
fDate
4/1/1997 12:00:00 AM
Firstpage
220
Lastpage
223
Abstract
Comparisons of the 1 V Josephson-array voltage standard (JAVS) of the Bureau International des Poids et Mesures (BIPM) were made with that of the Swedish National Testing and Research Institute (SP), Boras, Sweden, in April 1996. The results agree within an overall uncertainty of less than 1 nV
Keywords
measurement errors; measurement standards; superconducting junction devices; voltage measurement; 1 V; BIPM; Bureau International des Poids et Mesures; Josephson-array voltage standard; Swedish National Testing and Research Institute; comparison; overall uncertainty; Electrical resistance measurement; Filters; Frequency; Laboratories; Measurement standards; Metrology; NIST; Switches; Testing; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.571817