• DocumentCode
    1455925
  • Title

    AC resistance bridge based on the cryogenic current comparator

  • Author

    Seppä, Heikki ; Satrapinski, Alexandre

  • Author_Institution
    VTT Automation Meas. Technol., Espo, Finland
  • Volume
    46
  • Issue
    2
  • fYear
    1997
  • fDate
    4/1/1997 12:00:00 AM
  • Firstpage
    463
  • Lastpage
    466
  • Abstract
    This paper describes a low-frequency ac resistance bridge based on a Cryogenic Current Comparator (CCC). The bridge is a main tool for resistance traceability of 10 kΩ, 100 Ω, and 1 Ω to the quantum Hall resistance (QHR). The bridge is typically operated with a PC generating a sinusoidal ac signal ranging from (0.1 to 2) Hz. Using a vector analyzer as a signal source the bridge is used to 1 kHz. Both sinusoidal and pseudo random signals have been used to compare the frequency dependence of standard resistors against the QHR. At present the bridge can be used to perform accurate calibrations to 10 Hz, at higher frequencies the self-resonance of the CCC increases the uncertainty
  • Keywords
    SQUIDs; bridge instruments; calibration; cryogenics; current comparators; electric resistance measurement; measurement errors; quantum Hall effect; resistors; 0.1 Hz to 1 kHz; 1 ohm to 10 kohm; AC resistance bridge; SQUIDs; calibrations; cryogenic current comparator; pseudo random signals; quantum Hall resistance; resistance traceability; self-resonance; standard resistors; uncertainty; vector analyzer; Bridges; Cryogenics; Detectors; Electrical resistance measurement; Frequency dependence; Frequency measurement; Low-frequency noise; Resistors; SQUIDs; Signal generators;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.571886
  • Filename
    571886