DocumentCode :
1457275
Title :
Planning for camera-guided robotic radiosurgery
Author :
Schweikard, Achim ; Bodduluri, Mohan ; Adler, John R.
Author_Institution :
Dept. of Comput. Sci., Tech. Univ. Munchen, Germany
Volume :
14
Issue :
6
fYear :
1998
fDate :
12/1/1998 12:00:00 AM
Firstpage :
951
Lastpage :
962
Abstract :
Radiosurgery is a treatment for brain tumours and other lesions which employs a moving beam of photon radiation. In robotic radiosurgery, the radiation source is moved by a six degree-of-freedom robotic arm. A treatment planning system for robotic radiosurgery was developed. This planning system combines collision avoidance techniques and geometric algorithms for finding appropriate robot paths and beam activation profiles. The following design goals characterize our application: 1) ability to generate conformal 3D dose distributions; 2) transparency of the planning process; and 3) full use of the motion flexibility provided by the robotic arm. The new planning system is in clinical use in several leading medical institutions. Results obtained from the studies suggest that highly conformal distributions can be generated in a very effective way
Keywords :
computer vision; manipulators; path planning; radiation therapy; spatial reasoning; surgery; tumours; 3D dose distributions; brain tumours; camera-guided control; collision avoidance; computer vision; geometric reasoning; manipulator; motion planning; photon radiation; radiation therapy; robotic radiosurgery; surgery; treatment planning; Biomedical imaging; Collimators; Computed tomography; Lesions; Linear accelerators; Machine vision; Medical robotics; Medical treatment; Motion planning; Robots;
fLanguage :
English
Journal_Title :
Robotics and Automation, IEEE Transactions on
Publisher :
ieee
ISSN :
1042-296X
Type :
jour
DOI :
10.1109/70.736778
Filename :
736778
Link To Document :
بازگشت