DocumentCode :
1457570
Title :
Shape calibration of a conformal ultrasound therapy array
Author :
McGough, Robert J. ; Cindric, Davor ; Samulski, Thaddeus V.
Author_Institution :
Dept. of Radiat. Oncology, Duke Univ. Med. Center, Durham, NC, USA
Volume :
48
Issue :
2
fYear :
2001
fDate :
3/1/2001 12:00:00 AM
Firstpage :
494
Lastpage :
505
Abstract :
A conformal ultrasound phased array prototype with 96 elements was previously calibrated for electronic steering and focusing in a water tank. The procedure for calibrating the shape of this 2D therapy array consists of two steps. First, a least squares triangulation algorithm determines the element coordinates from a 21×21 grid of time delays. The triangulation algorithm also requires temperature measurements to compensate for variations in the speed of sound. Second, a Rayleigh-Sommerfeld formulation of the acoustic radiation integral is aligned to a second grid of measured pressure amplitudes in a least squares sense. This shape calibration procedure, which is applicable to a wide variety of ultrasound phased arrays, was tested on a square array panel consisting of 7-×7-mm elements operating at 617 kHz. The simulated fields generated by an array of 96 equivalent elements are consistent with the measured data, even in the fine structure away from the primary focus and sidelobes. These two calibration steps are sufficient for the simulation model to predict successfully the pressure field generated by this conformal ultrasound phased array prototype.
Keywords :
biomedical transducers; biomedical ultrasonics; calibration; least squares approximations; radiation therapy; ultrasonic transducer arrays; 2D therapy array; Rayleigh-Sommerfeld formulation; acoustic radiation integral; conformal ultrasound phased array prototype; conformal ultrasound therapy array; least squares triangulation algorithm; pressure field; shape calibration; temperature measurements; ultrasound phased arrays; Calibration; Delay effects; Least squares methods; Medical treatment; Phased arrays; Predictive models; Prototypes; Shape; Ultrasonic imaging; Ultrasonic variables measurement; Acoustics; Algorithms; Biomedical Engineering; Computer Simulation; Humans; Least-Squares Analysis; Temperature; Ultrasonic Therapy;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.911732
Filename :
911732
Link To Document :