DocumentCode
1458061
Title
Precise Alignment of a Magnetic Sensor in a Coordinate Measuring Machine
Author
Husstedt, Hendrik ; Ausserlechner, Udo ; Kaltenbacher, Manfred
Author_Institution
Alps-Adriatic Univ. Klagenfurt, Karnten, Austria
Volume
10
Issue
5
fYear
2010
fDate
5/1/2010 12:00:00 AM
Firstpage
984
Lastpage
990
Abstract
In order to measure strongly inhomogeneous magnetic fields it is straight forward to mount a magnetic sensor in a coordinate measuring machine (CMM). A scan of the field provides the components of the field vector w.r.t. the coordinates of the CMM reference frame. However, assembly tolerances of the sensor impair the field measurement. This paper explains a calibration procedure how to determine the misalignment of the sensor w.r.t. the reference frame of the CMM. To this end, the sensor needs at least two and for high accuracy four sensor elements on a single die with negligible hysteresis and offset. The calibration procedure requires a reference magnetic field with certain symmetry properties. A magnetic circuit generating this field is shown. Inaccuracies of the sensor elements and a curvature of the die surface are investigated theoretically and exemplary and result in a maximum error smaller than ±0.027°.
Keywords
coordinate measuring machines; magnetic field measurement; magnetic sensors; coordinate measuring machine; field vector; inhomogeneous magnetic fields; magnetic sensor; maximum error; negligible hysteresis; reference magnetic field; Assembly; Calibration; Coordinate measuring machines; Magnetic analysis; Magnetic field measurement; Magnetic hysteresis; Magnetic sensors; Packaging machines; Sensor arrays; Test equipment; Alignment; Hall sensors; coordinate measuring machine (CMM); sensor array;
fLanguage
English
Journal_Title
Sensors Journal, IEEE
Publisher
ieee
ISSN
1530-437X
Type
jour
DOI
10.1109/JSEN.2009.2037235
Filename
5440035
Link To Document