• DocumentCode
    1459401
  • Title

    Coherent scatter implementation for SimSET

  • Author

    Kaplan, M.S. ; Harrison, R.L. ; Vannoy, S.D.

  • Author_Institution
    Washington Univ., Seattle, WA, USA
  • Volume
    45
  • Issue
    6
  • fYear
    1998
  • fDate
    12/1/1998 12:00:00 AM
  • Firstpage
    3064
  • Lastpage
    3068
  • Abstract
    At 140 keV, 3% of photon scatter interactions in human tissues are coherent scatter; at Tl-201 emission energies, this fraction increases to approximately 7%. However, since coherent scatter at these energies is sharply forward-peaked, it is often the dominant scatter interaction at small angles. SimSET (Simulation System for Emission Tomography), which previously modeled only photoelectric absorption and Compton scatter, has been extended to include coherent scatter. The current implementation uses form factor and anomalous scattering amplitude data from the Livermore Evaluated Photon Data Library. Interaction probability and angular distribution tables for several human tissues and common detector materials were calculated using the independent atoms approximation and human-tissue composition data from the ICRP Reference Man. These data were also used to generate new tables for photoelectric absorption and Compton scatter, significantly improving the accuracy of SimSET and extending its photon tracking capability to lower photon energy (from 50 keV to 1 keV). The form, content, and structure of the tables were carefully designed for efficient data storage, access, and use by the software. The derived data tables and implementation of coherent scatter were validated by comparing simulation results to published differential cross-section data
  • Keywords
    Compton effect; digital simulation; emission tomography; gamma-ray scattering; medical image processing; photoelectricity; software packages; 140 keV; 50 to 1 keV; Compton scatter; ICRP Reference Man; Livermore Evaluated Photon Data Library; SimSET; angular distribution tables; anomalous scattering amplitude data; coherent scatter implementation; common detector materials; differential cross-section data; form factor; human tissues; human-tissue composition data; interaction probability; medical diagnostic imaging; nuclear medicine; photoelectric absorption; Absorption; Biological materials; Composite materials; Detectors; Electromagnetic scattering; Humans; Libraries; Particle scattering; Probability; Tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.737665
  • Filename
    737665