    Superconducting receiver coils for sodium magnetic resonance imaging

  • Author

    Miller, J.R. ; Zhang, K. ; Ma, Q.Y. ; Mun, I.K. ; Jung, K.J. ; Katz, J. ; Face, D.W. ; Kountz, D.J.

  • Author_Institution
    Dept. of Electr. Eng., Columbia Univ., New York, NY, USA
  • Volume
    43
  • Issue
    12
  • fYear
    1996
  • Firstpage
    1197
  • Lastpage
    1199
  • Abstract
    Presents the results from sodium magnetic resonance imaging (MRI) experiments using high-temperature superconducting (HTS) receiver coils. Sodium imaging has been shown to have great potential for the assessment of cell integrity but suffers from substantially lower signal-to-noise ratio (SNR) than that of hydrogen imaging. The use of an HTS receiver coil was found to significantly increase the SNR relative to an equivalent copper receiver coil at room temperature. The SNR gains afforded by HTS coils can also be used to decrease the imaging time.
  • Keywords
    biomedical NMR; biomedical equipment; sodium; superconducting coils; Cu; H; Na; cell integrity assessment; equivalent copper receiver coil; imaging time decrease; medical diagnostic imaging; medical instrumentation; room temperature; signal-to-noise ratio; sodium magnetic resonance imaging; superconducting receiver coils; Biomedical imaging; Copper; High temperature superconductors; Hydrogen; Low-frequency noise; Magnetic noise; Magnetic resonance imaging; Signal to noise ratio; Superconducting coils; Superconducting device noise; Copper; Electric Conductivity; Equipment Design; Fourier Analysis; Humans; Magnetic Resonance Imaging; Phantoms, Imaging; Reproducibility of Results; Time Factors;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.544344
  • Filename
    544344