DocumentCode
1462217
Title
Superconducting receiver coils for sodium magnetic resonance imaging
Author
Miller, J.R. ; Zhang, K. ; Ma, Q.Y. ; Mun, I.K. ; Jung, K.J. ; Katz, J. ; Face, D.W. ; Kountz, D.J.
Author_Institution
Dept. of Electr. Eng., Columbia Univ., New York, NY, USA
Volume
43
Issue
12
fYear
1996
Firstpage
1197
Lastpage
1199
Abstract
Presents the results from sodium magnetic resonance imaging (MRI) experiments using high-temperature superconducting (HTS) receiver coils. Sodium imaging has been shown to have great potential for the assessment of cell integrity but suffers from substantially lower signal-to-noise ratio (SNR) than that of hydrogen imaging. The use of an HTS receiver coil was found to significantly increase the SNR relative to an equivalent copper receiver coil at room temperature. The SNR gains afforded by HTS coils can also be used to decrease the imaging time.
Keywords
biomedical NMR; biomedical equipment; sodium; superconducting coils; Cu; H; Na; cell integrity assessment; equivalent copper receiver coil; imaging time decrease; medical diagnostic imaging; medical instrumentation; room temperature; signal-to-noise ratio; sodium magnetic resonance imaging; superconducting receiver coils; Biomedical imaging; Copper; High temperature superconductors; Hydrogen; Low-frequency noise; Magnetic noise; Magnetic resonance imaging; Signal to noise ratio; Superconducting coils; Superconducting device noise; Copper; Electric Conductivity; Equipment Design; Fourier Analysis; Humans; Magnetic Resonance Imaging; Phantoms, Imaging; Reproducibility of Results; Time Factors;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.544344
Filename
544344
Link To Document