DocumentCode
1462312
Title
A Monte Carlo optical code for the study of gamma cameras based on a scintillator read out by a silicon photodetector array
Author
Fiorini, Carlo
Author_Institution
Dipt. di Ingegneria Nucl., Politecnico di Milano, Italy
Volume
47
Issue
6
fYear
2000
fDate
12/1/2000 12:00:00 AM
Firstpage
1911
Lastpage
1917
Abstract
A Monte Carlo optical code for the optimization of scintillation gamma cameras based on a single scintillator coupled to arrays of Si photodiodes (PDs) or Si drift detectors (SDDs) have been developed. The simulator allows to calculate the expected gain (e-/keV), position resolution and energy resolution of the gamma detector for different geometries and boundary conditions at the crystal interfaces. A validation of the algorithm has been carried out by comparing the results of the simulations with the corresponding experimental results obtained on existing prototypes of detectors recently developed. An example of application of the proposed code in the study of a small scintillation camera is also reported
Keywords
Monte Carlo methods; gamma-ray detection; photodiodes; position sensitive particle detectors; silicon radiation detectors; solid scintillation detectors; Monte Carlo optical code; Si; Si drift detectors; Si photodiodes; energy resolution; gain; gamma cameras; position resolution; scintillation gamma cameras; Cameras; Energy resolution; Gamma ray detection; Gamma ray detectors; Monte Carlo methods; Optical arrays; Optical coupling; Photodiodes; Sensor arrays; Solid scintillation detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.914469
Filename
914469
Link To Document