    Precision AC–DC Difference Measurement System Based on a Programmable Josephson Voltage Standard

  • Author

    Budovsky, Ilya ; Georgakopoulos, Dimitrios ; Hagen, Thomas ; Sasaki, Hitoshi ; Yamamori, Hirotake

  • Author_Institution
    Nat. Meas. Inst., Lindfield, NSW, Australia
  • Volume
    60
  • Issue
    7
  • fYear
    2011
  • fDate
    7/1/2011 12:00:00 AM
  • Firstpage
    2439
  • Lastpage
    2444
  • Abstract
    A system designed for precision measurement of ac-dc difference of thermal voltage converters in terms of a programmable Josephson voltage standard is under development at the National Measurement Institute, Australia, in collaboration with the National Institute of Advanced Industrial Science and Technology, Japan. The agreement between the new system and calculable thermal converters is presently better than 0.5 μV/V at frequencies up to 40 Hz and voltages up to 0.75 V.
  • Keywords
    AC-DC power convertors; Josephson effect; programmable circuits; voltage measurement; IEEE thermal voltage converters; National Institute of Advanced Industrial Science and Technology; National Measurement Institute; frequency 40 Hz; precision AC-DC difference measurement system; programmable Josephson voltage standard; voltage 0.75 V; Current measurement; Radio frequency; Standards; Switches; Transient analysis; Uncertainty; Voltage measurement; Alternating-current–direct-current (ac–dc) transfer standards; Josephson arrays; Josephson devices; alternating voltage measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2011.2108554
  • Filename
    5722030