• DocumentCode
    1466855
  • Title

    Characterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes

  • Author

    Yoshida, Ken ; Stein, Richard B.

  • Author_Institution
    Div. of Neurosci., Alberta Univ., Edmonton, Alta., Canada
  • Volume
    46
  • Issue
    2
  • fYear
    1999
  • Firstpage
    226
  • Lastpage
    234
  • Abstract
    Longitudinal intrafascicular electrodes (LIFEs) are fine electrodes threaded into the extracellular space between axons in peripheral nerves or spinal roots. The authors are developing these electrodes for application in functional electrical stimulation and in basic physiology. An area of concern in chronic recording application of LIFEs is the possibility of electromyogram and other external noise sources masking the recorded neural signals. The authors characterized neural signals recorded by LIFEs and confirmed by three independent methods that increasing interelectrode spacing for bipolar LIFEs increases signal amplitude. The spectrum of neural signal from bipolar and monopolar LIFE lies between 300 Hz and 10 kHz. The amplitude of the spectrum increases with increasing interelectrode spacing, although the distribution is not affected. Single unit analysis of LIFE recordings show that they record selectively from units closest to the electrode active site. Units with conduction velocities ranging from 50-120 m/s were identified. Extraneural noise, as stimulus artifact or electromyogram, is much reduced with bipolar LIFE recording, as compared to monopolar recordings. Relative improvement in neural signal to extraneural noise increases with interelectrode spacing up to about 2 mm. Since there is no further improvement beyond 2 mm, the authors conclude that the preferred interelectrode spacing for bipolar LIFEs is 2 mm.
  • Keywords
    biomedical electrodes; cellular biophysics; electromyography; medical signal processing; neurophysiology; 2 mm; 300 Hz to 10 kHz; basic physiology; bipolar longitudinal intrafascicular electrodes; chronic recording application; conduction velocity; electrode active site; electromyogram; external noise sources; functional electrical stimulation; interelectrode spacing; noise rejection; peripheral nerves; signals characterization; single unit analysis; spinal roots; Distortion; Electrodes; Electromyography; Extracellular; Filtering; Muscles; Nerve fibers; Neuromuscular stimulation; Neuroscience; Physiology; Animals; Artifacts; Cats; Electrodes, Implanted; Electromyography; Hindlimb; Iridium; Muscle, Skeletal; Neurons; Platinum; Polytetrafluoroethylene; Sciatic Nerve; Signal Processing, Computer-Assisted; Spinal Nerve Roots;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.740885
  • Filename
    740885