DocumentCode
1467715
Title
Speckle tracking methods for ultrasonic elasticity imaging using short-time correlation
Author
Lubinski, Mark A. ; Emelianov, Stanislav Y. ; O´Donnell, Matthew
Author_Institution
Dept. of Biomed. Eng. & Electr. Eng., Michigan Univ., Ann Arbor, MI, USA
Volume
46
Issue
1
fYear
1999
Firstpage
82
Lastpage
96
Abstract
In ultrasound elasticity imaging, strain decorrelation is a major source of error in displacements estimated using correlation techniques. This error can be significantly decreased by reducing the correlation kernel. Additional gains in signal-to-noise ratio (SNR) are possible by filtering the correlation functions prior to displacement estimation. Tradeoffs between spatial resolution and estimate variance are discussed, and estimation in elasticity imaging is compared to traditional time-delay estimation. Simulations and experiments on gel-based phantoms are presented. The results demonstrate that high resolution, high SNR strain estimates can be computed using small correlation kernels (on the order of the autocorrelation width of the ultrasound signal) and correlation filtering.
Keywords
acoustic correlation; biological tissues; biomechanics; biomedical ultrasonics; elasticity; maximum likelihood estimation; medical image processing; speckle; autocorrelation width; biomedical US; correlation kernel; displacement estimation; elastography; gel-based phantoms; short-time correlation; signal-to-noise ratio; speckle tracking methods; strain decorrelation; ultrasonic elasticity imaging; ultrasound signal; Capacitive sensors; Decorrelation; Elasticity; Filtering; High-resolution imaging; Kernel; Signal to noise ratio; Spatial resolution; Speckle; Ultrasonic imaging;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.741427
Filename
741427
Link To Document