DocumentCode :
1469315
Title :
Spectral Characterization of the HIRDLS Flight Instrument From Prelaunch Calibration Data
Author :
Eden, Thomas, Jr. ; Barnett, John ; Gille, John ; Djotni, Karim ; Hepplewhite, Chris ; Oduleye, Olusoji ; Whitney, John
Author_Institution :
Nat. Center for Atmos. Res., Boulder, CO, USA
Volume :
48
Issue :
7
fYear :
2010
fDate :
7/1/2010 12:00:00 AM
Firstpage :
3028
Lastpage :
3038
Abstract :
Results from an instrument-level spectral characterization of the 21-channel High Resolution Dynamics Limb Sounder (HIRDLS) flight instrument will be presented. These data were obtained during the prelaunch calibration of HIRDLS at the University of Oxford (fall 2002). A monochromator, equipped with a controllable diffraction grating, was used to produce monochromatic light for these tests. The monochromator was housed, along with HIRDLS, in a large vacuum chamber. The monochromator was also equipped with a polarizer, which allowed data to be acquired at known polarizations for each channel. A calibration detector, with a flat spectral response, was used to measure the output from the monochromator. This paper will document the analysis procedures used to obtain a measured instrument spectral response for each channel, along with the associated error analyses for these measurements.
Keywords :
diffraction gratings; geophysical techniques; remote sensing; 21-channel high resolution dynamics limb sounder flight instrument; HIRDLS flight instrument; controllable diffraction grating; monochromator; prelaunch calibration data; Diffraction grating; monochromator; polarization; spectral;
fLanguage :
English
Journal_Title :
Geoscience and Remote Sensing, IEEE Transactions on
Publisher :
ieee
ISSN :
0196-2892
Type :
jour
DOI :
10.1109/TGRS.2010.2043676
Filename :
5446426
Link To Document :
بازگشت