• DocumentCode
    1470992
  • Title

    International comparison for RF power in the frequency range up to 18 GHz

  • Author

    De Vreede, Jan P M ; Korfage, Willemien ; Persson, Patrik ; Brunetti, Luciano ; Lopez, Valentin ; Petras, Ivan ; Morard, Pierre ; Hejsek, Frantisek ; Török, András ; Rühaak, Jürgen ; Ascroft, John ; Dressler, Erik ; Celep, Murat ; Lapuh, Rado ; Achkar, Jo

  • Author_Institution
    NMi Van Swinden Lab., Delft, Netherlands
  • Volume
    50
  • Issue
    2
  • fYear
    2001
  • fDate
    4/1/2001 12:00:00 AM
  • Firstpage
    409
  • Lastpage
    413
  • Abstract
    A Euromet international laboratory comparison (Project 393) has been carried out between 14 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The comparison is carried out using the normal equipment of the laboratory for high-quality external calibration. The results show good agreement in measuring the calibration coefficient of the thermistor mounts within the claimed expanded uncertainty (typically between 1% and 2%). It confirms the equivalence of national standards for RF power up to 18 GHz. In one case, corrective action is proposed
  • Keywords
    calibration; laboratories; measurement standards; measurement uncertainty; microwave measurement; power measurement; 18 GHz; Euromet international laboratory comparison; Project 393; RF power; Thermistor mounts; calibration coefficient; expanded uncertainty; high-quality external calibration; national standards; national standards laboratories; type N connectors; up to 18 GHz; Africa; Calibration; Connectors; Hafnium; Laboratories; Measurement standards; Metrology; Power measurement; Radio frequency; Thermistors;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.918154
  • Filename
    918154