International comparison of a pulse calibrator used in high-voltage impulse calibration
Author
Li, Yi ; Rungis, Juris ; McComb, Terry R. ; Lucas, Wolfgang
Author_Institution
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
Volume
50
Issue
2
fYear
2001
fDate
4/1/2001 12:00:00 AM
Firstpage
430
Lastpage
435
Abstract
Measurements on a pulse generator have been made at the NRC in Canada and the PTB in Germany. Most of the measured peak values (Vp ) among the laboratories agree to ±0.5%, the front times (T 1) to ±2.0%, and the times to half values (T2 ) to ±1.0%. The differences between calculated calibrator outputs and measured averages were found to be 0.13%, 0.63%, and 0.32% for Vp, T1, and T2, respectively. Contributions of impulse analysis software to the measurement errors are also discussed
Keywords
calibration; high-voltage techniques; impedance matching; measurement errors; measurement standards; measurement uncertainty; pulse generators; pulse measurement; voltage measurement; calculated calibrator outputs; high-voltage impulse calibration; impulse analysis software; impulse measurement; international comparison; low source impedance; matched calibrator; measured averages; measurement errors; measurement uncertainty; peak voltage measurement; pulse calibrator; pulse generator; Australia; Calibration; Circuit simulation; Councils; Laboratories; Pulse generation; Pulse measurements; Resistors; Switches; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.918159
Filename
918159
Link To Document