DocumentCode :
1471041
Title :
International comparison of a resistive divider at 100 kV DC
Author :
Li, Yi ; Rungis, Juris ; Jing, Tao ; Su, Tsung-han ; Chen, I.P. ; Lee, Dennis ; Shimizu, K.
Author_Institution :
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
Volume :
50
Issue :
2
fYear :
2001
fDate :
4/1/2001 12:00:00 AM
Firstpage :
436
Lastpage :
439
Abstract :
Values of the voltage divider ratio of a DC high-voltage divider, built and characterized by the NML of Australia, were determined at the Productivity and Standards Board of Singapore, the Center for Measurement Standards of Taiwan, the Standards and Calibration Laboratory of Hong Kong, and Japan Electric Meters Inspection Corporation by the comparison method, under an APMP program. The resistance values of the NML divider were also determined in the participating laboratories. Good agreement was found between the measured average resistance ratio and the measured average voltage ratio. The relative voltage dependence of the dividers has also been established
Keywords :
calibration; high-voltage techniques; thin film resistors; transfer standards; voltage dividers; voltage measurement; 100 kV; DC high-voltage divider; HV standards; average resistance ratio; average voltage ratio; calibration; international comparison; relative voltage dependence; resistive divider; thin film resistors; travelling divider; voltage divider ratio; Australia; Calibration; Electric resistance; Electrical resistance measurement; Laboratories; Measurement standards; Productivity; Standards Board; Voltage; Watthour meters;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.918160
Filename :
918160
Link To Document :