• DocumentCode
    1473357
  • Title

    Comments on "Is accurate recording of the ECG surface Laplacian feasible?" [with reply]

  • Author

    Jie Lian ; Srinivasa, Siddhartha ; Wo Chie Tsai ; Bin He ; Geselowitz, David B. ; Ferrara, James

  • Author_Institution
    Dept. of Bioeng., Illinois Univ., Chicago, IL, USA
  • Volume
    48
  • Issue
    5
  • fYear
    2001
  • fDate
    5/1/2001 12:00:00 AM
  • Firstpage
    610
  • Lastpage
    613
  • Abstract
    The authors comment on the work of D.B. Geselowitz and J. Ferrara (ibid., vol. 47, p. 1138, 2000) who concluded that the signal-to-noise ratio (SNR) of the ECG surface Laplacian (SL) is very low. Averaged over all recording channels on the anterolateral chest and averaged over all six subjects, when the noise level is estimated during the window of the P-R segment, our experimental results indicated that the potential ECG, regular Laplacian ECG, and diagonal Laplacian ECG have SNR of 105.70/spl plusmn/22.49, 33.73/spl plusmn/9.86, and 46.83/spl plusmn/12.84, respectively. This is in contrast with the magnitude of SNR reported in the text of the above paper, which was too low becaase of their choice of a residue window containing noise as well as signal. Geselowitz and Ferrara reply to these comments.
  • Keywords
    Laplace transforms; electrocardiography; medical signal processing; ECG surface Laplacian; accurate recording feasibility; anterolateral chest; body surface potential mapping; low signal-to-noise ratio; residue window; Biomedical engineering; Electrocardiography; Helium; Humans; Laplace equations; Noise level; Protocols; Shoulder; Signal to noise ratio; Voltage; Body Surface Potential Mapping; Humans; Male; Models, Cardiovascular; Models, Statistical; Reference Values; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.918603
  • Filename