Title :
Comments on "Is accurate recording of the ECG surface Laplacian feasible?" [with reply]
Author :
Jie Lian ; Srinivasa, Siddhartha ; Wo Chie Tsai ; Bin He ; Geselowitz, David B. ; Ferrara, James
Author_Institution :
Dept. of Bioeng., Illinois Univ., Chicago, IL, USA
fDate :
5/1/2001 12:00:00 AM
Abstract :
The authors comment on the work of D.B. Geselowitz and J. Ferrara (ibid., vol. 47, p. 1138, 2000) who concluded that the signal-to-noise ratio (SNR) of the ECG surface Laplacian (SL) is very low. Averaged over all recording channels on the anterolateral chest and averaged over all six subjects, when the noise level is estimated during the window of the P-R segment, our experimental results indicated that the potential ECG, regular Laplacian ECG, and diagonal Laplacian ECG have SNR of 105.70/spl plusmn/22.49, 33.73/spl plusmn/9.86, and 46.83/spl plusmn/12.84, respectively. This is in contrast with the magnitude of SNR reported in the text of the above paper, which was too low becaase of their choice of a residue window containing noise as well as signal. Geselowitz and Ferrara reply to these comments.
Keywords :
Laplace transforms; electrocardiography; medical signal processing; ECG surface Laplacian; accurate recording feasibility; anterolateral chest; body surface potential mapping; low signal-to-noise ratio; residue window; Biomedical engineering; Electrocardiography; Helium; Humans; Laplace equations; Noise level; Protocols; Shoulder; Signal to noise ratio; Voltage; Body Surface Potential Mapping; Humans; Male; Models, Cardiovascular; Models, Statistical; Reference Values; Signal Processing, Computer-Assisted;
Journal_Title :
Biomedical Engineering, IEEE Transactions on