• DocumentCode
    1474353
  • Title

    Properties of iodine cells contaminated by the heteromolecule 127I129I [He-Ne laser stability]

  • Author

    Balling, Petr ; Blabla, Jan

  • Author_Institution
    Lab. of Fundamental Metrol., Czech Metrol. Inst., Czech Republic
  • Volume
    47
  • Issue
    6
  • fYear
    1998
  • fDate
    12/1/1998 12:00:00 AM
  • Firstpage
    1481
  • Lastpage
    1486
  • Abstract
    The influence of contamination of the iodine cell by the heteromolecule is discussed. The results of the numerical model are compared with measured values for both the third (THT) and fifth (FHT) harmonic method of detection. Contamination causes a substantially lower frequency shift in the case of the FHT
  • Keywords
    gas lasers; harmonics; helium; iodine; laser frequency stability; neon; He-Ne; I2; contamination; fifth harmonic method; frequency shift; gas lasers; heteromolecule; iodine cells; numerical model; third harmonic method; Contamination; Filling; Frequency measurement; Hafnium; Impurities; Laser noise; Laser stability; Noise shaping; Numerical models; Pollution measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.746717
  • Filename
    746717