• DocumentCode
    1474754
  • Title

    Noise and spatial resolution in SQUID microscopy

  • Author

    Chatraphorn, S. ; Fleet, E.F. ; Wellstood, F.C. ; Knauss, L.A.

  • Author_Institution
    Dept. of Phys., Maryland Univ., College Park, MD, USA
  • Volume
    11
  • Issue
    1
  • fYear
    2001
  • fDate
    3/1/2001 12:00:00 AM
  • Firstpage
    234
  • Lastpage
    237
  • Abstract
    We have used a scanning SQUID microscope to image magnetic field generated by currents in integrated circuits. To obtain current paths in these circuits, we apply a magnetic inversion technique to the magnetic field data. We find that the spatial resolution obtained from this technique is related to the signal-to-noise ratio, the SQUID-sample separation and the data sampling interval. We describe in detail a mathematical model of how these parameters relate to the spatial resolution. Finally, we discuss the limitations of our apparatus, and how to achieve higher spatial resolution
  • Keywords
    SQUID magnetometers; integrated circuit testing; scanning probe microscopy; superconducting device noise; SQUID microscopy; SQUID-sample separation; current paths; currents; data sampling interval; integrated circuits; magnetic field; magnetic inversion technique; noise; scanning SQUID microscope; signal-to-noise ratio; spatial resolution; Image generation; Integrated circuit noise; Magnetic circuits; Magnetic fields; Magnetic force microscopy; Magnetic noise; Magnetic separation; SQUIDs; Signal to noise ratio; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/77.919327
  • Filename
    919327