DocumentCode
1475074
Title
An Active Quench Protection System for MRI Magnets
Author
Huang, Xianrui ; Wu, Anbo B. ; Xu, Minfeng ; Chu, Xu ; Yang, Chao ; Zhao, Yan
Author_Institution
GE China Technol. Center, Shanghai, China
Volume
20
Issue
3
fYear
2010
fDate
6/1/2010 12:00:00 AM
Firstpage
2091
Lastpage
2094
Abstract
An active quench protection system is developed to safely protect a 1.5 T superconducting MRI magnet. The protection system includes a digital quench detection controller, a set of quench heaters and voltage taps, a battery backed DC power supply and solid-state switches. The quench controller takes the voltage signals from the voltage taps over superconducting coils and monitor a quench onset. To distinguish a quench signal from operational signals, the quench controller uses a pre-determined moving average voltage over 50 ms as the threshold signal for quench detection. The threshold signal is more than 10 times larger than any of the normal operation signals to avoid false quenching. Once detected moving average voltage exceeds the threshold, the quench controller will close solid-state relays and power quench heaters to accelerate the quench propagation over the entire magnet. The system has been tested in a MRI magnet test module.
Keywords
biomedical MRI; superconducting magnets; DC power supply; active quench protection; digital quench detection controller; magnetic flux density 1.5 T; power quench heaters; solid-state relays; solid-state switches; superconducting MRI magnet; Active quench protection; moving averaged voltage; quench; superconducting magnets;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/TASC.2010.2044033
Filename
5451140
Link To Document