• DocumentCode
    1478654
  • Title

    Mean blood velocity measurement with a narrow ultrasound beam and an asymmetric velocity profile

  • Author

    Willink, Robin

  • Author_Institution
    Ind. Res. Ltd., Lower Hutt, New Zealand
  • Volume
    46
  • Issue
    3
  • fYear
    1999
  • fDate
    3/1/1999 12:00:00 AM
  • Firstpage
    362
  • Lastpage
    364
  • Abstract
    Previous discussions of the measurement of spatial mean blood velocity using Doppler ultrasound with a narrow beam have required the velocity profile in the cross section to be symmetric about the vessel axis. A type of asymmetry is put forward which incurs no error in mean velocity measurement when the beam is directed through the paint of maximum velocity. Assuming correct alignment of the beam, errors due to profile asymmetry are, therefore, related to the degree of departure of the asymmetry from this acceptable form.
  • Keywords
    Doppler measurement; biomedical ultrasonics; blood flow measurement; measurement errors; Doppler ultrasound; acoustic beams; asymmetric velocity profile; biomedical acoustics; mean blood velocity measurement; narrow ultrasound beam; profile asymmetry errors; spatial mean blood velocity; Acoustic beams; Acoustic measurements; Biomedical acoustics; Biomedical measurements; Blood flow; Blood vessels; Error correction; Ultrasonic imaging; Ultrasonic variables measurement; Velocity measurement; Blood Flow Velocity; Blood Vessels; Models, Cardiovascular; Pulsatile Flow; Transducers; Ultrasonography, Doppler;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.748989
  • Filename