• DocumentCode
    1482966
  • Title

    Split-type magnetic field generator with wide homogeneous region

  • Author

    Shirai, Terumitsu

  • Author_Institution
    Japan Electr. Meters Inspection Corp., Tokyo, Japan
  • Volume
    37
  • Issue
    3
  • fYear
    2001
  • fDate
    5/1/2001 12:00:00 AM
  • Firstpage
    1091
  • Lastpage
    1094
  • Abstract
    This paper discusses the homogeneity of a magnetic field generator for measurement of nuclear magnetic resonance (NMR). It shows that inhomogeneity is caused by an excess number of turns and a discrepancy from the design dimensions. A control circuit consisting of a pair of shim coils and a few Hall elements can improve the homogeneity. With it, a homogeneity of the magnetic field in the split coil of better than ±0.05% over 50 mm was obtained
  • Keywords
    magnetic field measurement; magnetometers; nuclear magnetic resonance; solenoids; Hall elements; homogeneous region; magnetic field generator; nuclear magnetic resonance; shim coils; split-type magnetic field generator; Coils; Magnetic field measurement; Magnetic fields; Magnetic resonance; Nonuniform electric fields; Nuclear magnetic resonance; Nuclear measurements; Nuclear power generation; Paramagnetic resonance; Solenoids;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.920481
  • Filename
    920481