Title
Split-type magnetic field generator with wide homogeneous region
Author
Shirai, Terumitsu
Author_Institution
Japan Electr. Meters Inspection Corp., Tokyo, Japan
Volume
37
Issue
3
fYear
2001
fDate
5/1/2001 12:00:00 AM
Firstpage
1091
Lastpage
1094
Abstract
This paper discusses the homogeneity of a magnetic field generator for measurement of nuclear magnetic resonance (NMR). It shows that inhomogeneity is caused by an excess number of turns and a discrepancy from the design dimensions. A control circuit consisting of a pair of shim coils and a few Hall elements can improve the homogeneity. With it, a homogeneity of the magnetic field in the split coil of better than ±0.05% over 50 mm was obtained
Keywords
magnetic field measurement; magnetometers; nuclear magnetic resonance; solenoids; Hall elements; homogeneous region; magnetic field generator; nuclear magnetic resonance; shim coils; split-type magnetic field generator; Coils; Magnetic field measurement; Magnetic fields; Magnetic resonance; Nonuniform electric fields; Nuclear magnetic resonance; Nuclear measurements; Nuclear power generation; Paramagnetic resonance; Solenoids;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.920481
Filename