On Super-Resolution With an Experimental Microwave Tomography System
Author :
Gilmore, Colin ; Mojabi, Puyan ; Zakaria, Amer ; Pistorius, Stephen ; LoVetri, Joe
Author_Institution :
Med. Phys., CancerCare Manitoba, Winnipeg, MB, Canada
fDate :
7/2/1905 12:00:00 AM
Abstract :
The resolution of an experimental microwave tomography (MWT) system is investigated. Using two cylindrical nylon targets and an operating frequency of 5 GHz, a separation resolution of 2 mm, or 1/30 of a wavelength, is achieved. While this resolution is among the highest reported in the literature, it is not a sufficiently robust indicator of the expected resolution obtainable for complex targets, and this is shown with further examples of more complicated targets. However, the basic separation resolution limit obtained is a good way of comparing various aspects of different MWT systems.
Keywords :
image resolution; medical image processing; microwave imaging; tomography; cylindrical nylon targets; frequency 5 GHz; microwave tomography system; separation resolution; separation resolution limit; Electromagnetic tomography; image resolution; microwave imaging;