DocumentCode
1483406
Title
The BIPM Compact Josephson Voltage Standard
Author
Solve, Stéphane ; Chayramy, Régis
Author_Institution
Int. Bur. of Weights & Measures (BIPM), Sevres, France
Volume
60
Issue
7
fYear
2011
fDate
7/1/2011 12:00:00 AM
Firstpage
2366
Lastpage
2371
Abstract
Fifteen National Metrology Institutes have recently confirmed their interest in participating in the International Bureau of Weights and Measures (BIPM) direct on-site comparison of Josephson array voltage standards at the level of 10 V. This comparison has been ongoing since 1994, and the BIPM primary standard involved needs to be replaced by a new standard built from components based on modern technology. We present here in detail the components of the new BIPM Josephson voltage standard. The main goal during its development was to realize a standard that is more easily transportable, more convenient to use, and as efficient and reliable as its predecessor. Both primary standards have already shown agreement within a total combined relative uncertainty of 1.4 × 10-10 (k = 1).
Keywords
Josephson effect; measurement standards; voltage measurement; Josephson array voltage standards; Arrays; Batteries; Frequency measurement; Lead; Radio frequency; Uncertainty; Voltage measurement; International Bureau of Weights and Measures (BIPM) direct on-site comparison; Josephson voltage standard (JVS);
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2011.2108610