• DocumentCode
    1488598
  • Title

    The coil design of the superconducting MRI magnet

  • Author

    Fujita, M.

  • Author_Institution
    Fuji Electr. Corp. Res. & Dev. Ltd., Chiba, Japan
  • Volume
    24
  • Issue
    6
  • fYear
    1988
  • fDate
    11/1/1988 12:00:00 AM
  • Firstpage
    2907
  • Lastpage
    2909
  • Abstract
    A program has been developed to optimize the coil configuration of highly homogeneous magnetic resonance imaging (MRI) magnets with primary and canceling coils. To reduce the amount of leakage fluxes generated by the magnet, the ampere-turn ratio for the primary and canceling coils is set to balance their total magnetic moment. To realize a highly homogeneous magnetic field over the imaging space, the configuration of the main coils is designed to compensate for the flux density components up to the tenth order of the inhomogeneous magnetic field induced by the canceling coils
  • Keywords
    CAD; biomedical NMR; coils; medical computing; superconducting magnets; ampere-turn ratio; canceling coils; coil configuration optimisation; coil design; compensation up to tenth order; highly homogeneous magnetic field; inhomogeneous magnetic field; leakage flux reduction; magnetic resonance imaging; primary coils; superconducting MRI magnet; total magnetic moment balance; Gaussian processes; Iron; Magnetic fields; Magnetic flux; Magnetic moments; Magnetic resonance imaging; Magnetic shielding; Research and development; Superconducting coils; Superconducting magnets;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.92284
  • Filename