DocumentCode
1490695
Title
Design of 2.92-mm Coaxial Adiabatic Lines for Quasi-Ideal Twin Microcalorimeter
Author
Vremera, Emil T. ; Brunetti, Luciano
Author_Institution
Fac. of Electr. Eng., Gheorghe Asachi Tech. Univ. of Iasi, Iasi, Romania
Volume
61
Issue
6
fYear
2012
fDate
6/1/2012 12:00:00 AM
Firstpage
1692
Lastpage
1702
Abstract
This paper discusses high-frequency power standard realization by means of broadband microcalorimetry improvement by using the concept of a true-twin microcalorimeter. An enhanced design for the twin microcalorimeter allows application of new measurement methods. The adiabatic line, as a key part in defining the twin-microcalorimeter behavior, is investigated from thermal and electrical points of view by means of thermal-to-electrical models, by simulation, and by both thermal and electrical measurements. Furthermore, the pairing quality of these parts is evaluated, for getting a quasi-ideal microcalorimeter with two identical channels.
Keywords
calorimeters; coaxial cables; coaxial adiabatic line; electrical measurement; power standard; quasi-ideal twin microcalorimeter; size 2.92 mm; thermal measurement; thermal to electrical model; Conductors; Temperature measurement; Thermal conductivity; Thermal resistance; Thermal sensors; $S$ -parameters measurement; Adiabatic line; feeding path; microcalorimeter; thermal model; thermal simulation;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2012.2190331
Filename
6180217