• DocumentCode
    1490695
  • Title

    Design of 2.92-mm Coaxial Adiabatic Lines for Quasi-Ideal Twin Microcalorimeter

  • Author

    Vremera, Emil T. ; Brunetti, Luciano

  • Author_Institution
    Fac. of Electr. Eng., Gheorghe Asachi Tech. Univ. of Iasi, Iasi, Romania
  • Volume
    61
  • Issue
    6
  • fYear
    2012
  • fDate
    6/1/2012 12:00:00 AM
  • Firstpage
    1692
  • Lastpage
    1702
  • Abstract
    This paper discusses high-frequency power standard realization by means of broadband microcalorimetry improvement by using the concept of a true-twin microcalorimeter. An enhanced design for the twin microcalorimeter allows application of new measurement methods. The adiabatic line, as a key part in defining the twin-microcalorimeter behavior, is investigated from thermal and electrical points of view by means of thermal-to-electrical models, by simulation, and by both thermal and electrical measurements. Furthermore, the pairing quality of these parts is evaluated, for getting a quasi-ideal microcalorimeter with two identical channels.
  • Keywords
    calorimeters; coaxial cables; coaxial adiabatic line; electrical measurement; power standard; quasi-ideal twin microcalorimeter; size 2.92 mm; thermal measurement; thermal to electrical model; Conductors; Temperature measurement; Thermal conductivity; Thermal resistance; Thermal sensors; $S$-parameters measurement; Adiabatic line; feeding path; microcalorimeter; thermal model; thermal simulation;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2012.2190331
  • Filename
    6180217