• DocumentCode
    1492998
  • Title

    Error in the thermoresistive sensor-based feedback measurement configuration due to DC amplifier offset voltage

  • Author

    Freire, Raimundo Carlos Silvério ; Deep, Gurdip Singh

  • Author_Institution
    Dept. of Electr. Eng., Univ. Fed. da Paraiba, Joao Pessoa, Brazil
  • Volume
    48
  • Issue
    1
  • fYear
    1999
  • fDate
    2/1/1999 12:00:00 AM
  • Firstpage
    50
  • Lastpage
    54
  • Abstract
    Feedback circuits with thermoresistive sensors have been widely used in the measurement of thermal radiation and fluid flow. In general, the operation of these circuits is explained, assuming an ideal operational amplifier used in the feedback configuration. The effect of the operational amplifier offset voltage on the static performance has been analyzed. The error introduced when used as a measuring instrument, due to input offset voltage, is estimated
  • Keywords
    DC amplifiers; anemometers; circuit feedback; electric sensing devices; flow measurement; measurement errors; operational amplifiers; radiometry; thermoelectric devices; DC amplifier offset voltage; anemometers; error estimation; feedback configuration; feedback measurement; fluid flow; measurement of thermal radiation; operational amplifier; operational amplifier offset voltage; radiation detector; static performance; thermoresistive sensors; Bandwidth; Feedback circuits; Fluid flow; Fluid flow measurement; Operational amplifiers; Sensor phenomena and characterization; Temperature sensors; Thermal sensors; Thermoresistivity; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.755059
  • Filename
    755059