Title
Design and construction of the phantom prostate for the treatment in conformational radiotherapy
Author
Yaya Castaneda, Guillermo ; Castaneda, B.
Author_Institution
Dipt. de Ing., Pontificia Univ. Catolica del Peru (PUCP), Lima, Peru
Volume
10
Issue
2
fYear
2012
fDate
3/1/2012 12:00:00 AM
Firstpage
1452
Lastpage
1458
Abstract
Phantoms made of synthetic materials are commonly used in conformational radiotherapy to verify the proper operation of therapeutic medical equipment. The main objective of the paper is to design and build a prostate phantom to be used in the calibration of planning systems. The proposed phantom is made of nylon and acrylic. These inexpensive materials are readily available in Peru and have comparable density to soft tissue and bone. The phantom was used in the calibration of a planning system. Final results showed an average error of 5.49% between the measurements made on the phantom and planned values. These results are in the range of uncertainty recommended by the International Commission on Radiation Units and Measurements (ICRU).
Keywords
bone; calibration; phantoms; polymers; radiation therapy; International Commission on Radiation Units and Measurements; calibration; conformational radiotherapy; planning systems; prostate phantom; synthetic materials; Calibration; Detectors; Laboratories; Materials; Neodymium; Phantoms; Visualization; CT; Ionizing radiation detectors; Phantom; conformational radiotherapy; linear accelerator clinical;
fLanguage
English
Journal_Title
Latin America Transactions, IEEE (Revista IEEE America Latina)
Publisher
ieee
ISSN
1548-0992
Type
jour
DOI
10.1109/TLA.2012.6187586
Filename
6187586
Link To Document