DocumentCode :
1503138
Title :
Calibration of Rogowski Coils With an Integrator at High Currents
Author :
Draxler, Karel ; Styblikova, Renata ; Hlavacek, Jan ; Prochazka, Radek
Author_Institution :
Dept. of Meas., Czech Tech. Univ. in Prague, Prague, Czech Republic
Volume :
60
Issue :
7
fYear :
2011
fDate :
7/1/2011 12:00:00 AM
Firstpage :
2434
Lastpage :
2438
Abstract :
This paper describes the calibration of Rogowski coils with an integrator for current measurement in the current range of up to 30 kA, at mains frequency. A current loop of ten turns with a current of 3 kA was used for calibration. The magnetic field generated in this way differs from the field of one turn placed at the center and passed with a current of 30 kA. The influence of the unevenness of winding and the cross section of the Rogowski coil may cause a calibration error when using a current loop. The difference in the results between using a current loop and one turn, and the measurement uncertainties are given here. Calculations of the mutual inductances of typical Rogowski coils and their uncertainties are presented in the Appendix.
Keywords :
calibration; coils; electric current measurement; integrating circuits; measurement uncertainty; Rogowski coils; calibration error; current 3 kA; current 30 kA; current loop; current measurement; integrator; measurement uncertainty; winding unevenness; Calibration; Coils; Conductors; Current measurement; Inductance; Uncertainty; Voltage measurement; Calibration; Rogowski coil (Rc); cross-section winding; high currents; winding unevenness;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2011.2126090