DocumentCode :
1505380
Title :
Measurement of functional residual capacity through the transient phase of He dilution in newborns
Author :
Bouferrache, Belkacem ; Krim, Gerard ; Léké, André ; Freville, Michel ; Libert, Jean-Pierre
Author_Institution :
Sch. of Med., URAPC, Amiens, France
Volume :
48
Issue :
7
fYear :
2001
fDate :
7/1/2001 12:00:00 AM
Firstpage :
834
Lastpage :
838
Abstract :
Helium dilution maneuver is used to determine the functional residual capacity (FRC) 14 newborns ages 1-5 mo. The model equation describes the changing alveolar fractions of He and the ventilation promoted by a rebreathing procedure that does not exceed 40 s. The model does not involve the volume of the rebreathing bag usually needed when applying rebreathing technique and which is a source of error. The equation is discretized and solved for recorded data obtained with equipment adapted to newborns. Results show a strong relationship between FRC and the biometrical indexes, and confirm these found in the literature featuring that the measurement duration of FRC can be considerably shortened.
Keywords :
biodiffusion; biomedical measurement; helium; paediatrics; physiological models; pneumodynamics; 40 s; He; He dilution transient phase; alveolar fractions; biometrical indexes; error source; functional residual capacity measurement; gas exchanges modeling; measurement duration; newborns; rebreathing bag volume; Asphyxia; Circuits; Equations; Gases; Helium; Lungs; Pediatrics; Phase measurement; Ventilation; Volume measurement; Body Surface Area; Data Collection; Functional Residual Capacity; Helium; Humans; Infant; Infant, Newborn; Models, Biological; Pulmonary Gas Exchange; Spirometry;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.930909
Filename :
930909
Link To Document :
بازگشت