DocumentCode :
1505545
Title :
A moving-actuator type electromechanical total artificial heart. I. Linear type and mock circulation experiments
Author :
Min, Byoung G. ; Kim, Hee C. ; Lee, Sang H. ; Kim, Jong W. ; Kim, Jin T. ; Kim, In Y. ; Kim, Sung W. ; Diegel, Paul D. ; Olsen, Don B.
Author_Institution :
Dept. of Biomed. Eng., Seoul Nat. Univ., South Korea
Volume :
37
Issue :
12
fYear :
1990
Firstpage :
1186
Lastpage :
1194
Abstract :
A novel type of motor-driven total artificial heart system with a moving-actuator mechanism has been developed. The prototype system consists of a brushless DC motor inside a rolling cylinder, two arcshaped pusher plates, and two polyurethane sacs. The moving-actuator-type electromechanical pump has the structural advantages of small size and light weight, as compared to other motor-driven pumps with fixed actuator mechanisms. Results of mock circulation tests with a cardiac output of 9 L/min at an aortic pressure of 120 mm Hg and a heart rate of 120 b/m are reported.
Keywords :
artificial organs; cardiology; 120 mmHg; aortic pressure; arcshaped pusher plates; brushless DC motor; cardiac output; heart rate; mock circulation experiments; motor-driven pumps; moving-actuator type electromechanical total artificial heart; polyurethane sacs; rolling cylinder; Artificial heart; Bridges; Brushless DC motors; Couplings; Heart rate; Humans; Prototypes; Testing; Vacuum systems; Wires; Algorithms; Animals; Blood Circulation; Heart, Artificial; Humans; Models, Cardiovascular; Prosthesis Design;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.64461
Filename :
64461
Link To Document :
بازگشت