• DocumentCode
    1507365
  • Title

    A Stimulator ASIC Featuring Versatile Management for Vestibular Prostheses

  • Author

    Dai Jiang ; Demosthenous, A. ; Perkins, T. ; Xiao Liu ; Donaldson, N.

  • Author_Institution
    Dept. of Electron. & Electr. Eng., Univ. Coll. London, London, UK
  • Volume
    5
  • Issue
    2
  • fYear
    2011
  • fDate
    4/1/2011 12:00:00 AM
  • Firstpage
    147
  • Lastpage
    159
  • Abstract
    This paper presents a multichannel stimulator ASIC for an implantable vestibular prosthesis. The system features versatile stimulation management which allows fine setting of the parameters for biphasic stimulation pulses. To address the problem of charge imbalance due to rounding errors, the digital processor can calculate and provide accurate charge correction. A technique to reduce the data rate to the stimulator is described. The stimulator ASIC was implemented in 0.6-μ m high-voltage CMOS technology occupying an area of 2.27 mm2. The measured performance of the ASIC has been verified using vestibular electrodes in saline.
  • Keywords
    CMOS integrated circuits; application specific integrated circuits; biomedical electronics; prosthetics; biphasic stimulation pulses; charge correction; digital processor; high-voltage CMOS technology; implantable vestibular prosthesis; multichannel stimulator ASIC; saline; size 0.6 mum; versatile stimulation management; vestibular electrodes; Application specific integrated circuits; Clocks; Electrodes; Irrigation; Prosthetics; Radiation detectors; Synchronization; Charge balance; data compression; implanted device; neural stimulation; stimulation management; vestibular prosthesis;
  • fLanguage
    English
  • Journal_Title
    Biomedical Circuits and Systems, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1932-4545
  • Type

    jour

  • DOI
    10.1109/TBCAS.2011.2138139
  • Filename
    5759288