DocumentCode
1508681
Title
Comparison between a magneto-optical method and Fannin´s technique for the measurement of Brown´s relaxation frequency of ferrofluids
Author
Payet, B. ; Siblini, A. ; Blanc-Mignon, M.F. ; Noyel, G.
Author_Institution
L.R.C.C.I., St.-Etienne Univ., France
Volume
35
Issue
3
fYear
1999
fDate
5/1/1999 12:00:00 AM
Firstpage
2018
Lastpage
2023
Abstract
A magneto-optical method, conceived in our laboratory to measure Brown´s relaxation frequency of ferrofluids, is compared with Fannin´s technique based on impedance measurements (slit toroid). We have developed a complete theory and an experimental setup of the first method and have briefly described the second one. We present a study of Brown´s relaxation of cobalt ferrite Fe2CoO4 in different carrier liquids (glycerol, diethylene glycol, dibutyl phthalate, and water). The two methods show very close results, proving the high sensitivity of both. The relation between the viscosity η and Brown´s relaxation frequency fB has been verified. Moreover the analysis of relaxation and viscosity versus temperature has enabled us to estimate the hydrodynamic diameter of particles
Keywords
Brownian motion; cobalt compounds; ferrites; magnetic fluids; magnetic particles; magnetic permeability; magnetic relaxation; magnetic variables measurement; magneto-optical effects; particle size measurement; viscosity; Brown´s relaxation frequency; CoFe2O4; Fannin´s technique; Fe2CoO4; carrier liquids; ferrofluids; hydrodynamic diameter; impedance measurements; magneto-optical method; slit toroid; Cobalt; Ferrites; Frequency measurement; Impedance measurement; Iron; Laboratories; Liquids; Temperature sensors; Time of arrival estimation; Viscosity;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.764903
Filename
764903