DocumentCode :
1508956
Title :
Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry
Author :
Fetics, Barry ; Nevo, Erez ; Chen, Chen-Huan ; Kass, David A.
Author_Institution :
Dept. of Biomed. Eng., Johns Hopkins Med. Inst., Baltimore, MD, USA
Volume :
46
Issue :
6
fYear :
1999
fDate :
6/1/1999 12:00:00 AM
Firstpage :
698
Lastpage :
706
Abstract :
Aortic pressure can be estimated noninvasively by applying a transfer function (TF) to radial tonometry signals. This study compares the performance of prior approaches, based on Fourier transform and inverted aortic-to-radial model, with direct radial-to-aortic autoregressive exogenous (ARX) model. Simultaneous invasive aortic pressure and radial tonometry pressure were recorded during rest in 39 patients in the supine position. Individual radial-aortic TF´s were estimated from 20 patients, and the average TF was used to predict aortic pressures in the remaining 19 patients. The direct average TF yielded accurate aortic systolic pressure estimation (error 0.4±2.9 mmHg) and good reproduction of the aortic pressure waveform (root mean squared error 2.2±0.9 mmHg). The inverted reverse TF (aortic radial) yielded comparable results, while the Fourier-based TF had worse performance. Individual direct TF provided improved predictive accuracy only for indexes which are based on higher frequency components of the waveform (augmentation index, systolic time period). An ARX average TF can be used to accurately estimate central aortic pressure waveform parameters from noninvasive radial pulse tracings, and its performance is superior to previous techniques.
Keywords :
blood pressure measurement; medical signal processing; physiological models; transfer functions; aortic pressure; augmentation index; direct radial-to-aortic autoregressive exogenous model; higher frequency waveform components; inverted aortic-to-radial model; noninvasive estimation; noninvasive radial pulse tracings; parametric model derivation; radial tonometry; supine position; systolic time period; Biomedical measurements; Blood pressure; Cardiology; Fourier transforms; Frequency; Parametric statistics; Pulse measurements; Steady-state; Transfer functions; Transient analysis; Adult; Aged; Algorithms; Aorta; Artifacts; Bias (Epidemiology); Blood Pressure Determination; Calibration; Female; Fourier Analysis; Humans; Male; Middle Aged; Models, Cardiovascular; Predictive Value of Tests; Radial Artery; Reproducibility of Results; Signal Processing, Computer-Assisted; Supine Position; Systole;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.764946
Filename :
764946
Link To Document :
بازگشت