DocumentCode :
1509007
Title :
Wiener beamforming and the coherence factor in ultrasound imaging
Author :
Nilsen, Carl-Inge Colombo ; Holm, Sverre
Author_Institution :
Dept. of Inf., Univ. of Oslo, Oslo, Norway
Volume :
57
Issue :
6
fYear :
2010
fDate :
6/1/2010 12:00:00 AM
Firstpage :
1329
Lastpage :
1346
Abstract :
The coherence factor (CF) is used for aberration correction and sidelobe suppression in ultrasound imaging. Unfortunately, it suffers from artifacts when the SNR is low. We show how the CF can be interpreted as an implementation of the Wiener postfilter for a delay-and-sum beamformer. In addition, we show that a minimum-variance, distortionless-response beamformer followed by CF weighting can be interpreted as an implementation of the Wiener beamformer. These interpretations provide us with a theoretical framework for analyzing and improving CF-based methods. We use this theory to develop more robust implementations for both the Wiener postfilter and beamformer. The performance of these implementations is shown on simulated and real data.
Keywords :
Wiener filters; acoustic signal processing; ultrasonic imaging; CF weighting; Wiener beamforming; Wiener postfilter; aberration correction; coherence factor; delay-and-sum beamformer; distortionless-response beamformer; sidelobe suppression; signal-to-noise ratio; ultrasound imaging; Array signal processing; Delay; Robustness; Ultrasonic imaging; Algorithms; Computer Simulation; Cysts; Databases, Factual; Humans; Image Processing, Computer-Assisted; Models, Theoretical; Ultrasonography;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2010.1553
Filename :
5480176
Link To Document :
بازگشت